3458a_system_voltmeter

3458A Digital Multimeter

First Catalog 1989
Last Catalog ?
Class 8 1/2-digit

Introduced in 1989; see HP Journal of April 1989.

“An 8 1/2-Digit Digital Multimeter Capable of 100,000 Readings per Second and Two-Source Calibration. A highly linear and extremely flexible analog-to-digital converter and a state-of-the-art design give this DMM new performance and measurement capabilities for automated test, calibration laboratory, or R&D applications.”

The 3458A multimeter is HP-IB (IEEE 488) programmable and an enhanced implementation of a multislope ADC achieving a range of speeds and resolutions from 16-bit resolution at 100,000 readings per second to 28-bit resolution at six readings per second. The ADC linearity has deviations that are less than 0.1 ppm of input.

The 3458A was released under the Hewlett Packard brand, but also had Agilent branding and is currently selling under the Keysight brand. Agilent is a spinoff of the Test & Measurement division (including optics, instrumentation, chemical analysis, and medical equipment product lines) of HP, while Keysight is a further spinoff of the test and measurement product lines from Agilent.

digital_voltmeters_and_multimeters:3458a-front.jpg

Basic specifications

DCV

  • Ranges : 100mV to 1000V
  • Sensitivity: 10nV
  • Accuracy 1yr(of reading+range): 9+3ppm@100mV, 8+0.3@1V, 8+0.05ppm@10V, 10+0.3ppm@100V, 10+0.1@1000V
  • Accuracy 24h(of reading+range): 2.5+3ppm@100mV, 1.5+0.3@1V, 0.5+0.05ppm@10V, 2.5+0.3ppm@100V, 2.5+0.1@1000V
  • Input resistance : 10GΩ on ranges up to 10V; 10MΩ +- 1% on 100V and 1000V ranges

ACV

  • >2MHz true RMS
  • Ranges: 10mV to 1000V
  • Resolution: 10nV to 1mV (6.5digit)
  • Best Accuracy (%reading+%range): 0.02+0.01@100mV-10V (100Hz-20KHz)
  • Input impedance: 1MΩ with < 140pF

Resistance (2W/4W)

  • Ranges: 10Ω to 1GΩ
  • Sensitivity: 10 μΩ to 100Ω
  • Open Circuit voltage: 12V (10-10MΩ), 5V (100MΩ & 1GΩ)

DCI

  • Ranges: 100nA, 1μA, 10μA, 100μA, 1mA, 10mA, 100mA, 1A
  • Burden voltage : 0.05V, 0.045V. 0.05V, 0.075V, 0.1V, 0.1V, 0.25V, <1.5V

ACI

  • Ranges: 100μA, 1mA, 10mA, 100mA, 1A
  • Burden voltage : 0.1V, 0.1V,0.1V, 0.25V, <1.5V

Agilent 3458A Datasheet

Options/Plugins
Option 001 Extended Reading Memory (Expands total to 148 k bytes)
Option 002 High Stability (4 ppm/year) Reference
Option HFL Fluke version 3458A/HFL with 2ppm/year stability
Option 907 Front Handles Kit (5062-3988)
Option 908 Rack Mount Kit (5062-3974)
Option 909 Rack Mount Kit with handles (5062-3975)

Firmware

The 3458A contains one main CPU (Motorola 68HC000P8), one microcontroller in the analog domain (Intel 80C51BH) and one on the outguard power supply (Intel 80C51BH). The latest outguard power supply has changed.

Mods & repairs

Other software